Skip to content
Home » DASYLab Modules for sound level & sound power measurement

DASYLab Modules for sound level & sound power measurement

    (EN 60651 and EN 60804)

    ROGA  DASYLab 12

    • Time weighting: fast, slow, impulse, leq following DIN IEC 651 and DIN IEC 804.
    • Filters A, B and C following EN 60651 and EN 60804
    • Switchable 10 Hz High Pass filter
    • Comfortable calibration of measurement up to 16 input channels
    • Freely selectable calibration level

    Download Module Datasheet

    Sound Level Measurement Module
    The sound level measurement module has the following features:

      • Time weighting: fast, slow, impulse, leq following DIN IEC 651 and DIN IEC 804.
      • Easy microphone calibration with a pistonphone; in calibration mode, DasyLAB 6 detects the channel/microphone to calibrate and calculates the correct values.
      • The correction values are stored with the worksheet.
      • The module has 16 inputs and 16 outputs for the weightedand dB scaled sound levels.
      • We recommend a 16 bit AD-converter with microphonepower supply.
      • A hardware dB A weighting is nice to have, but not necessary.
      • Simultaneous Sample and Hold is absolutely necessary for sound pressure measurement.

    Sound Power Measurement Module
    The sound power measurement module can calculate sound power for a maximum of 16 input channels (from the sound level measurement module) 1 output channel, switchable to:

    • Measurement surface sound power.
    • Sound power level (SPL) of all active channels.The module properties allow you to set the four corrective values in dB or use the “wizards” to determine them:
      • K0: Correction value for air pressure and temperature. Direct input of the dB value or pressure and temperature. (Only necessary for class 1 measurements according to DIN 45 635.)
      • K1: Correction value for extraneous noise correction (back-ground noise, signal-to-noise-ratio). Direct input of thedB value or taken from last measurement.
      • K2: Correction value for environment feedback (reflections).Direct input of the dB value or input of the room’s metrics:– volume – reverberation time …
      • Ls: Correction value for for the enveloping surface. Direct input of the dB value or input of the surface metrics (guided with graphics). Like DIN (2a, 2c, b) – Spherical, hemisphere, quarter globe – Cuboids (detached, at a wall, at a wall and ceiling)

      The module works in accordance with the following standards:DIN 45 635, DIN EN 23 741, ISO 3741, DIN EN 23 742, ISO 3742, DIN EN 23 744, EN ISO 3744, DIN EN 21 680, ISO 6395

    (Source: Roga Instruments)